- name
CAPA system management for medical device QMS. Covers root cause analysis, corrective action planning, effectiveness verification, and CAPA metrics. Use when running CAPA investigations, 5-Why analysi
SKILL.md
The skill file stays in the publisher's repository.
Paste the copied command in a terminal. It installs this SKILL.md into that agent.
npx skills add alirezarezvani/claude-skills --skill capa-officer -a cursor -yPaste the copied command in a terminal. It installs this SKILL.md into that agent.
npx skills add alirezarezvani/claude-skills --skill capa-officer -a claude-code -yPaste the copied command in a terminal. It installs this SKILL.md into that agent.
npx skills add alirezarezvani/claude-skills --skill capa-officer -a codex -yPaste the copied command in a terminal. It installs this SKILL.md into that agent.
npx skills add alirezarezvani/claude-skills --skill capa-officer -a grok -ySend the copied message in a Grok Bot chat. The Bot fetches this SKILL.md from GitHub and adds it to the skills all your Bots share.
Install this skill: https://github.com/alirezarezvani/claude-skills/tree/HEAD/.gemini/skills/capa-officerFrom the publisher
Corrective and Preventive Action (CAPA) management within Quality Management Systems, focusing on systematic root cause analysis, action implementation, and effectiveness verification.
Conduct systematic CAPA investigation from initiation through closure:
Select and apply appropriate RCA methodology based on problem characteristics.
Is the issue safety-critical or involves system reliability?
├── Yes → Use FAULT TREE ANALYSIS
└── No → Is human error the suspected primary cause?
├── Yes → Use HUMAN FACTORS ANALYSIS
└── No → How many potential contributing factors?
├── 1-2 factors (linear causation) → Use 5 WHY ANALYSIS
├── 3-6 factors (complex, systemic) → Use FISHBONE DIAGRAM
└── Unknown/proactive assessment → Use FMEA
Use when: Single-cause issues with linear causation, process deviations with clear failure point.
Template:
PROBLEM: [Clear, specific statement]
WHY 1: Why did [problem] occur?
BECAUSE: [First-level cause]
EVIDENCE: [Supporting data]
WHY 2: Why did [first-level cause] occur?
BECAUSE: [Second-level cause]
EVIDENCE: [Supporting data]
WHY 3: Why did [second-level cause] occur?
BECAUSE: [Third-level cause]
EVIDENCE: [Supporting data]
WHY 4: Why did [third-level cause] occur?
BECAUSE: [Fourth-level cause]
EVIDENCE: [Supporting data]
WHY 5: Why did [fourth-level cause] occur?
BECAUSE: [Root cause]
EVIDENCE: [Supporting data]
Example - Calibration Overdue:
PROBLEM: pH meter (EQ-042) found 2 months overdue for calibration
WHY 1: Why was calibration overdue?
BECAUSE: Equipment was not on calibration schedule
EVIDENCE: Calibration schedule reviewed, EQ-042 not listed
WHY 2: Why was it not on the schedule?
BECAUSE: Schedule not updated when equipment was purchased
EVIDENCE: Purchase date 2023-06-15, schedule dated 2023-01-01
WHY 3: Why was the schedule not updated?
BECAUSE: No process requires schedule update at equipment purchase
EVIDENCE: SOP-EQ-001 reviewed, no such requirement
WHY 4: Why is there no such requirement?
BECAUSE: Procedure written before equipment tracking was centralized
EVIDENCE: SOP last revised 2019, equipment system implemented 2021
WHY 5: Why has procedure not been updated?
BECAUSE: Periodic review did not assess compatibility with new systems
EVIDENCE: No review against new equipment system documented
ROOT CAUSE: Procedure review process does not assess compatibility
with organizational systems implemented after original procedure creation.
See references/rca-methodologies.md for complete method details and templates.
Before proceeding to action planning, validate root cause:
Develop effective actions addressing identified root causes:
ACTION PLAN TEMPLATE
CAPA Number: [CAPA-XXXX]
Root Cause: [Identified root cause]
ACTION 1: [Specific action description]
- Type: [ ] Containment [ ] Correction [ ] Corrective [ ] Preventive
- Responsible: [Name, Title]
- Due Date: [YYYY-MM-DD]
- Resources: [Required resources]
- Success Criteria: [Measurable outcome]
- Verification Method: [How success will be verified]
ACTION 2: [Specific action description]
...
IMPLEMENTATION TIMELINE:
Week 1: [Milestone]
Week 2: [Milestone]
Week 4: [Milestone]
Week 8: [Milestone]
APPROVAL:
CAPA Owner: _____________ Date: _______
Process Owner: _____________ Date: _______
QA Manager: _____________ Date: _______
Verify corrective actions achieved intended results:
Did recurrence occur during verification period?
├── Yes → CAPA INEFFECTIVE (re-investigate root cause)
└── No → Were all effectiveness criteria met?
├── Yes → CAPA EFFECTIVE (proceed to closure)
└── No → Extent of gap?
├── Minor gap → Extend verification or accept with justification
└── Significant gap → CAPA INEFFECTIVE (revise actions)
See references/effectiveness-verification-guide.md for detailed procedures.
Monitor CAPA program performance through key indicators.
Monthly CAPA status report includes:
references/rca-methodologies.md contains:
references/effectiveness-verification-guide.md contains:
# Generate CAPA status report
python scripts/capa_tracker.py --capas capas.json
# Interactive mode for manual entry
python scripts/capa_tracker.py --interactive
# JSON output for integration
python scripts/capa_tracker.py --capas capas.json --output json
# Generate sample data file
python scripts/capa_tracker.py --sample > sample_capas.json
Calculates and reports:
{
"capas": [
{
"capa_number": "CAPA-2024-001",
"title": "Calibration overdue for pH meter",
"description": "pH meter EQ-042 found 2 months overdue",
"source": "AUDIT",
"severity": "MAJOR",
"status": "VERIFICATION",
"open_date": "2024-06-15",
"target_date": "2024-08-15",
"owner": "J. Smith",
"root_cause": "Procedure review gap",
"corrective_action": "Updated SOP-EQ-001"
}
]
}
⚠️ STATUS — QMSR transition (effective 2026-02-02): FDA's Quality Management System Regulation (QMSR) final rule (89 FR 7496) amended 21 CFR Part 820 to incorporate ISO 13485:2016 by reference and removed the legacy QSR subsection structure. The CAPA section number 820.100 no longer exists in the CFR — it is retained below only as a familiar index. The current FDA authority for CAPA is ISO 13485:2016 §8.5.2 (corrective action) and §8.5.3 (preventive action) (see clause table above), with complaint-handling additions in retained 21 CFR 820.35. Cite the ISO 13485 clauses — not 820.100 — in current compliance documentation.
CAPA elements required under ISO 13485 §8.5.2/§8.5.3 (legacy QSR 820.100, historical):
Decision discipline: The tools in this skill structure investigations and track CAPA status — they do not certify CAPA closure or compliance. CAPA effectiveness conclusions and closure decisions are yours to make and must be reviewed and signed off by the named CAPA owner and Quality function; route regulatory-classification questions (e.g., reportability, 21 CFR 803 MDR, recall under 21 CFR 806) to Regulatory Affairs.
© 2026 Skills Rack.
---name: capa-officerdescription: CAPA system management for medical device QMS. Covers root cause analysis, corrective action planning…---# .gemini/skills/capa-officer/SKILL.mdOverview
CAPA system management for medical device QMS. Covers root cause analysis, corrective action planning, effectiveness verification, and CAPA metrics. Use when running CAPA investigations, 5-Why analysis, fishbone diagrams, root cause determination, corrective action tracking, effectiveness verification, or CAPA program optimization.
FAQ
Download .gemini/skills/capa-officer/SKILL.md from the publisher, or copy the path and open the repository. PlugMCP lists the skill and does not host the file.
The publisher keeps it at https://github.com/alirezarezvani/claude-skills.
Claude, Cursor, Codex, and Grok Build can read a SKILL.md you add from the publisher's repository. Grok Bot installs it when you send the skill's GitHub folder link in a chat.